8.24. FLUORIDE HIGH RANGE
SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENT
Code
HI93739A‑0
HI93739B‑0
REAGENT SETS
HI93739‑01
HI93739‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Fluoride HR
described in the Method Selection section
(see page 18).
• Add 2.00 mL of
Reagent A to the cuvette and use the pipette to fill
up the cuvette to the 10 mL mark with
Fluoride High Range Reagent B.
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
88
0.0 to 20.0 mg/L (as F
0.1 mg/L
±0.5 mg/L ±3% of reading at 25 °C
LED with narrow band interference filter @ 575 nm
Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18
edition, SPADNS method.
th
Description
Fluoride High Range Reagent A
Fluoride High Range Reagent B
Reagents for 100 tests
Reagents for 300 tests
method using the procedure
HI93739A‑0
Fluoride High Range
HI93739B‑0
)
‑
Quantity
2 mL
8 mL
2 mL
10 mL