• Remove the cuvette and add 5 drops of
pH Reagent Indicator. Replace the cap and mix the
solution.
• Reinsert the cuvette into the instrument and close the lid.
• Press the Read key to start the reading. The instrument displays the result in pH.
HI93710‑0