8.29. HARDNESS, TOTAL HIGH RANGE

SPECIFICATIONS
Range
400 to 750 mg/L (as CaCO
Resolution
1 mg/L
Accuracy
±10 mg/L ±2% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 466 nm
Method
Adaptation of the EPA recommended method 130.1.
REQUIRED REAGENTS
Code
Description
HI93735IND‑0
Hardness Indicator Reagent
HI93735A‑HR
Hardness High Range Reagent A
HI93735B‑0
Hardness Buffer Reagent B
HI93735C‑0
Fixing Reagent
REAGENT SETS
HI93735‑02
Reagents for 100 tests (HR, 400 to 750 mg/L)
HI93735‑0
Reagents for 300 tests (LR ‑ 100 tests, MR ‑ 100 tests, HR ‑ 100 tests)
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Total Hardness HR
described in the Method Selection section (see page 18).
• Add 0.5 mL of unreacted sample to the cuvette. Add
0.5 mL of
HI93735IND‑0
• With the plastic dropper fill the cuvette up to the 10 mL
mark with
HI93735A‑HR
• Add two drops of
HI93735B‑0
Reagent B. Replace the cap and invert 5 times to mix.
)
3
method using the procedure
Hardness Indicator Reagent.
Hardness High Range Reagent A.
Hardness Buffer
Quantity
0.5 mL
9 mL
2 drops
1 packet
103