Hanna Instruments HI83300 Instruction Manual page 130

Multiparameter
Table of Contents

Advertisement

Remove the cuvette and add one packet of
Nickel High Range Reagent. Replace the cap and shake
gently until completely dissolved.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 1 minute and press Read. When the timer ends the meter will perform the reading. The
instrument displays the concentration in g/L of nickel (Ni).
INTERFERENCES
Interference may be caused by copper.