• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 3 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in µg/L of phosphorus (P).
• Press  or  to access the second level functions.
• Press the Chem Frm key to convert the result to µg/L of phosphate (PO
pentoxide (P
O
).
2
5
• Press  or  to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 50 mg/L
Silica above 50 mg/L
Silicate above 10 mg/L
Copper above 10 mg/L
Hydrogen sulfide, arsenate, turbid sample and highly buffered samples also interfere.