• Press Read to start the reading. The instrument displays the results in mg/L of nickel (Ni).
Note: A temperature above 30 °C may cause turbidity. In this case add 2-3 drops of
HI93703-51
zeroing the meter and reading the sample.
INTERFERENCES
Interference may be caused by:
Co
must not be present
2+
Fe
must not be present
2+
Al
above 32 mg/L
3+
Ca
above 1000 mg/L (as CaCO
2+
Cd
above 20 mg/L
2+
Cl¯ above 8000 mg/L
Cr
above 20 mg/L
3+
Cr
above 40 mg/L
6+
Cu
above 15 mg/L
2+
F¯ above 20 mg/L
Fe
above 10 mg/L
3+
K
above 500 mg/L
+
Mg
above 400 mg/L
2+
Mn
above 25 mg/L
2+
Mo
above 60 mg/L
6+
Na
above 5000 mg/L
+
Pb
above 20 mg/L
2+
Zn
above 30 mg/L
2+
128
Dispersing Agent to each cuvette and swirl until turbidity is removed before
)