Nitrate - Hanna Instruments HI83300 Instruction Manual

Multiparameter
Table of Contents

Advertisement

8.40. NITRATE

SPECIFICATIONS
Range
0.0 to 30.0 mg/L (as NO
Resolution
0.1 mg/L
Accuracy
±0.5 mg/L ±10% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 525 nm
Method
Adaptation of the cadmium reduction method.
REQUIRED REAGENTS
Code
Description
HI93728‑0
Nitrate Reagent
REAGENT SETS
HI93728‑01
Reagents for 100 tests
HI93728‑03
Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Nitrate
method using the procedure described in
the Method Selection section (see page 18).
• Fill the cuvette with 10 mL of sample, (up to the mark),
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
‑ N)
3
Quantity
1 packet
10 mL
131