• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
POWDER REAGENT PROCEDURE
• Add the content of one packet of
Reagent. Replace the cap and shake gently for 20 seconds.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or alternatively,
wait for 1 minute and press Read. When the timer ends the meter will perform the reading. The
instrument displays the results in mg/L of chlorine (Cl
LIQUID REAGENT PROCEDURE
• To an empty cuvette add 3 drops of
Chlorine Reagent A and 3 drops of
Chlorine Reagent B.
HI93701‑0
Free Chlorine
).
2
HI93701A‑F
Free
HI93701B‑F
Free
63