3. SPECIFICATIONS

Measurement Channels
Range
Resolution
Accuracy
Light Source
Bandpass Filter Bandwidth
Absorbance
Bandpass Filter Wavelength Accuracy ±1.0 nm
Light Detector
Cuvette Types
Number of Methods
Range
Resolution
Accuracy
pH
Temperature Compensation
Calibration
Electrode
Range
Temperature
Resolution
Accuracy
Logging
Display
USB‑A (Host) Functions
USB‑B (Device) Functions
Battery Life
Additional
Specifications
Power Supply
Environment
Dimensions
Weight
*Limits will be reduced to actual probe/sensor limits.
5 x optical channels;
1 x digital electrode channel (pH measurement)
0.000 to 4.000 Abs
0.001 Abs
±0.003 Abs (at 1.000 Abs)
light emitting diode
8 nm
silicon photocell
round, 24.6 mm diameter
65
‑2.00 to 16.00 pH (± 1000.0 mV)*
0.01 pH (0.1 mV)
±0.01 pH (±0.2 mV) (@ 25 °C / 77 °F)
ATC (‑5.0 to 100.0 °C; 23.0 to 212.0 °F)*
2 points, eligible from 5 available buffers (4.01,
6.86, 7.01, 9.18, 10.01 pH)
Intelligent pH / temperature electrode
‑20.0 to 120.0°C (‑4.0 to 248.0 °F)
0.1 °C (0.1 °F)
±0.5 °C (±0.9 °F ) (@ 25 °C / 77 °F)
1000 readings (mixed photometer and electrode)
128 x 64 pixel B/W LCD with backlight
mass‑storage host
power input, mass‑storage device
> 500 photometer measurements, or 50 hours of
continuous pH measurement
5 Vdc USB 2.0 power adapter/type micro‑B connector
3.7 Vdc Li‑polymer rechargeable battery,
non‑serviceable
0 to 50 °C (32 to 122 °F);
0 to 95% RH, non‑serviceable
206 x 177 x 97 mm (8.1 x 7.0 x 3.8")
1.0 kg (2.2 lbs.)