Hanna Instruments HI83300 Instruction Manual page 73

Multiparameter
Table of Contents

Advertisement

• Add one packet of
HI95771A‑0
Range Reagent A and one packet
Chlorine Ultra High Range Reagent B. Replace the cap and
shake gently for 20 seconds.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 1 minute and press Read. When the timer ends the meter will perform the reading. The
instrument displays the results in mg/L of chlorine (Cl
INTERFERENCES
Interference may be caused by:
Bromine (Br
), Oxidized Manganese, Chromium, Chlorine Dioxide (ClO
2
Total Chlorine Ultra High
HI95771B‑0
Total
).
2
), Ozone (O
) and Iodine (I
2
3
).
2
73