Ammonia High Range - Hanna Instruments HI83300 Instruction Manual

Multiparameter
Table of Contents

Advertisement

8.6. AMMONIA HIGH RANGE

SPECIFICATIONS
Range
0.0 to 100.0 mg/L (as NH
Resolution
0.1 mg/L
Accuracy
±0.5 mg/L ±5% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 420 nm
Method
Adaptation of the ASTM Manual of Water and Environmental Technology,
D1426, Nessler method.
REQUIRED REAGENTS
Code
Description
HI93733A‑0
Ammonia High Range Reagent A
HI93733B‑0
Ammonia High Range Reagent B
REAGENT SETS
HI93733‑01
Reagents for 100 tests
HI93733‑03
Reagents for 300 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Ammonia HR
Method Selection section (see page 18).
• Add 1mL of unreacted sample to the cuvette using 1mL
syringe.
• Use the pipette to fill the cuvette up to the 10 mL mark
with
HI93733B‑0
Ammonia High Range Reagent B.
Replace the cap and mix the solution.
• Place the cuvette into the holder and close the lid.
‑N)
3
method using the procedure described in the
Quantity
4 drops
9 mL
1 mL
45