8.34. MAGNESIUM

SPECIFICATIONS
Range
0 to 150 mg/L (as Mg
Resolution
1 mg/L
Accuracy
±5 mg/L ±3% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 466 nm
Method
Adaptation of the Calmagite method.
REQUIRED REAGENTS
Code
Description
HI93752A‑Mg
Magnesium Reagent A
HI93752B‑Mg
Magnesium Reagent B
REAGENT SETS
HI937520‑01
Reagents for 50 tests
HI937520‑03
Reagents for 150 tests
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Magnesium
method using the procedure
described in the Method Selection section (see page 18).
• Add 1 mL of
HI93752A‑Mg
cuvette using a 1 mL syringe and use the pipette to fill
the cuvette up to the 10 mL mark with the
Magnesium Reagent B.
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
)
2+
Magnesium Reagent A to the
HI93752B‑Mg
Quantity
1 mL
9 mL
115