8.31. IODINE

SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI93718‑0
REAGENT SETS
HI93718‑01
HI93718‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Iodine
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
108
0.0 to 12.5 mg/L (as I
0.1 mg/L
±0.1 mg/L ±5% of reading at 25 °C
LED with narrow band interference filter @ 525 nm
Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, DPD method.
Description
Iodine Reagent
Reagents for 100 tests
Reagents for 300 tests
method using the procedure described in the
)
2
Quantity
1 packet
10 mL