Hanna Instruments HI83300 Instruction Manual page 66

Multiparameter
Table of Contents

Advertisement

• Add one packet of
Replace the cap and shake gently for 20 seconds.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 1 minute and press Read. When the timer ends the meter will perform the reading. The
instrument displays the results in mg/L of chlorine (Cl
INTERFERENCES
Interference may be caused by:
Alkalinity: above 1,000 mg/L CaCO
25 mg/L CaCO
3
develop the full amount of color or it may rapidly fade (negative error). To resolve this, neutralize
the sample with diluted HCl.
Acidity: above 150 mg/L CaCO
fade (negative error). To resolve this, neutralize the sample with diluted NaOH.
Hardness: in case of water with hardness greater than 500 mg/L CaCO
approximately 2 minutes after adding the powder reagent.
Bromine (Br
), Chloride dioxide (ClO
2
(O
): positive error.
3
66
HI95762‑0
Free Chlorine Reagent.
if present as bicarbonate (HCO
3
if present as carbonate (CO
. May not reliably develop the full amount of color or it may rapidly
3
), Iodine (I
2
).
2
sample pH < 8.3); above
3
, sample pH > 9.0). In both cases, it will not reliably
2‑
3
), Oxidized Manganese and Chromium, Ozone
2
, shake the sample for
3