8.38. NICKEL LOW RANGE

SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI93740A‑0
HI93740B‑0
HI93740C‑0
HI93703‑51
REAGENT SETS
HI93740‑01
HI93740‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Nickel LR
Method Selection section (see page 18).
Note: For best results samples should be between 20 and 24 °C.
• Fill one graduated beaker with 25 mL of deionized water (blank)
and another one with 25 mL of sample.
• Add one packet of
beaker. Cap and swirl gently until the reagent is dissolved.
Note: If sample contains iron (Fe
powder is dissolved before continuing.
• Add 1 mL of
beaker, and swirl to mix.
126
0.000 to 1.000 mg/L (as Ni)
0.001 mg/L
±0.010 mg/L ±7% of reading at 25 °C
LED with narrow band interference filter @ 575 nm
Adaptation of the PAN method.
Description
Nickel Low Range Reagent A
Nickel Low Range Reagent B
Nickel Low Range Reagent C
Dispersing Agent (optional reagent)
Reagents for 50 tests
Reagents for 150 tests
method using the procedure described in the
HI93740A‑0
Nickel Low Range Reagent A to each
), it is important that all
3+
HI93740B‑0
Nickel Low Range Reagent B to each
Quantity
2 packets
2 mL
2 packets
4‑6 drops