8.7. BROMINE
SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI93716‑0
REAGENT SETS
HI93716‑01
HI93716‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Bromine
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
48
0.00 to 8.00 mg/L (as Br
0.01 mg/L
±0.08 mg/L ±3% of reading at 25 °C
LED with narrow band interference filter @ 525 nm
Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18
edition, DPD method.
th
Description
Quantity
Bromine Reagent
1 packet
Reagents for 100 tests
Reagents for 300 tests
method using the procedure described in the
)
2
10 mL