For more information about Hanna Instruments and our products, visit www.hannainst.com or e-mail us at sales@hannainst.com. For technical support, contact your local Hanna Instruments office or e-mail us at tech@hannainst.com.
Remove the instrument and accessories from the packaging and examine it carefully. For further assistance, please contact your local Hanna Instruments office or email us at tech@hannainst.com.
Each instrument is delivered in a rugged case and is supplied with:
Note: Save all packing material until you are sure that the meter works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.
HI83141-1 is a portable, water-resistant meter designed to be paired with HI1230B combination pH electrode and HI7662 temperature probe for pH, mV, and temperature measurements. Optionally, the meter can be paired with HI3131B refillable ORP electrode (not supplied with the product) for accurate ORP measurements.
HI1230B pH
HI7662 Temperature
Meter
pH | |
Range * | 0.00 to 14.00 pH |
Resolution | 0.01 pH |
Accuracy | ±0.01 pH (@25°C / 77°F) |
Temperature compensation | Automatic, from 0 to 100°C (32 to 212°F) |
Calibration | Manual, 2-point (trimmers) Offset: ±1.00 pH; Slope: 80 to 110% |
mV | |
Range * | ±1999 mV |
Resolution | 1 mV |
Accuracy | ±1 mV (@25°C / 77°F) |
Temperature | |
Range* | 0.0 to 100.0°C (32.0 to 212.0°F) |
Resolution | 0.1°C / 0.1°F |
Accuracy | ±0.4°C / ±0.8°F (excluding probe error) |
Probes (included) | HI1230B pH electrode; HI7662 temperature |
Battery type | 1 x 9V Alkaline |
Battery life | Approximately 150 hours of continuous use |
Environment | 0 to 50°C (32 to 122°F); 100% RH |
Auto shut-off | After 8 minutes of non-use |
Dimensions | 145 x 80 x 36 mm (5.7 x 3.1 x 1.4'') |
Weight | 230 g (8.1 oz.) |
* The range may be limited by the probe's limits.
HI1230B pH Probe
Reference | double, Ag/AgCl |
Junction | ceramic, single |
Electrolyte | gel |
Recommended Operating Temp. | –5 to 70°C (23 to 158°F) |
Max. Pressure | 2 bar |
Range | 0 to 12 pH |
Tip / Shape | spheric, Ø 7.5 mm (0.29") |
Glass type | LT (low temperature) |
Body material | PEI |
Dimensions | Length: 163.5 mm (6.43") Shaft length: 120 mm (4.72") Shaft diameter: 12 mm (0.47") |
Cable length | 1 m (3.3'); coaxial |
HI7662 Temperature Probe
Dimensions | Total length: 173.5 mm (6.8") Active part: 105 mm (4.1"); Ø 3.6 mm (1.4") |
Cable length | 1 m (3.3') |
Note: If the "BATT" tag is displayed blinking, battery level is too low and the battery needs to be replaced.
Align the HI1230B BNC connector with the socket and plug it in.
Rotate the locking shell until it clicks in place.
Plug the HI7662 probe RCA connector into the designated socket.
Press the On/Off key to turn the meter on. Initialization screen briefly displays all the LCD segments followed by the battery status, alerting user to the remaining battery life.
For high accuracy, frequent calibrations are recommended.
Additionally, the pH range should be recalibrated:
One- or two-point calibration can be performed using one of the following standard buffer solutions: pH 4.01, 6.86 (NIST), 7.01, 9.18 (NIST), or pH 10.01.
When a two-point calibration is required, use pH 7.01 or pH 6.86 buffer as first calibration point.
Use pH 7.01 (HI7007) (or pH 6.86 NIST equivalent) for neutral samples, pH 4.01 (HI7004) for acidic samples, pH 10.01 (HI7010) (or pH 9.18 NIST equivalent) for alkaline samples.
For best results, use a rinse beaker and a separate calibration beaker for each buffer. Discard rinsing buffers after use.
Note: If measurements are taken in different samples successively, rinse the probes between measurements to avoid cross contamination. After cleaning, rinse the probes with some of the samples to be measured next.
If measured value is outside the parameter limit of the probe, the maximum or minimum value is displayed blinking.
Temp | pH Values | ||||
°C | °F | 4.01 | 7.01 | 10.01 | |
0 | 32 | 4.01 | 7.13 | 10.32 | |
5 | 41 | 4.00 | 7.10 | 10.24 | |
10 | 50 | 4.00 | 7.07 | 10.18 | |
15 | 59 | 4.00 | 7.04 | 10.12 | |
20 | 68 | 4.00 | 7.03 | 10.06 | |
25 | 77 | 4.01 | 7.01 | 10.01 | |
30 | 86 | 4.02 | 7.00 | 9.96 | |
35 | 95 | 4.03 | 6.99 | 9.92 | |
40 | 104 | 4.04 | 6.98 | 9.88 | |
45 | 113 | 4.05 | 6.98 | 9.85 | |
50 | 122 | 4.06 | 6.98 | 9.82 | |
55 | 131 | 4.07 | 6.98 | 9.79 | |
60 | 140 | 4.09 | 6.98 | 9.77 | |
65 | 149 | 4.11 | 6.99 | 9.76 | |
70 | 158 | 4.12 | 6.99 | 9.75 | |
75 | 167 | 4.14 | 7.00 | 9.74 | |
80 | 176 | 4.16 | 7.01 | 9.73 | |
85 | 185 | 4.17 | 7.02 | 9.74 | |
90 | 194 | 4.19 | 7.03 | 9.75 | |
95 | 203 | 4.20 | 7.04 | 9.76 |
Ordering info. Product description | |||||
HI1230B | Plastic body, double-junction, gel, combination pH electrode with BNC connector and 1 m (3.28 ft) cable | ||||
HI3131B | Glass-body, refillable, combination platinum ORP electrode with BNC connector and 1 m (3.28 ft) cable | ||||
HI7662 | Temperature probe with 1 m (3.28 ft) cable | ||||
HI7004M | pH 4.01 buffer solution, 230 mL | ||||
HI7006M | pH 6.86 buffer solution, 250 ml | ||||
HI7007M | pH 7.01 buffer solution, 230 mL | ||||
HI7009M | pH 9.18 buffer solution, 250 ml | ||||
HI7010M | pH 10.01 buffer solution, 230 mL | ||||
HI70300M | Storage solution, 230 mL | ||||
HI7061M | General cleaning solution, 230 mL | ||||
HI7091L | Reducing pretreatment solution, 500 mL + 14 g (set) | ||||
HI7092M | Oxidizing pretreatment solution, 250 mL | ||||
HI731326 | Calibration screwdriver (20 pcs.) | ||||
HI76405 | Electrode holder |
All Hanna® instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment.
The product should not be treated as household waste. Instead, hand it over to the appropriate collection point for the recycling of electrical and electronic equipment, which will conserve natural resources.
Before using this meter, make sure that it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the meter's performance. For your and the meter's safety do not use or store the meter in hazardous environments.
HI83141-1 is warranted for a period of two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Electrodes and probes are warranted for a period of six months.
If service is required, contact your local Hanna Instruments® office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments office, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction, or appearance of its products without advance notice.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Hanna Instruments HI83141-1 - Portable PH/mV/Temperature Meter Manual
Advertisement
Need help?
Do you have a question about the HI83141-1 and is the answer not in the manual?
Questions and answers