6.2.13.
Oil flow
The function is only available if the chosen fuel is an oil.
Calling up the function:
>
[
Performing the measurement:
1. Select the parameters
Pressure
2. Enter values. [▲],
-
The calculated oil burner capacity (in kW) is displayed.
Options:
>
[Options]
>
[Options]
is deleted.
>
[Options]
>
[Options]
(kg/h > gal/h
6.2.14.
Ambient CO
This function is not available when the
selected.
✓ An ambient CO probe (recommended) or a flue gas probe must
be connected.
Call up function:
>
[
[OK].
→
]
Measurements
(no effect on calculation): [▲],
[▼]
→ Clipboard: Data are saved to the clipboard
→
Delete
→ Save: The readings are saved in a protocol.
→
Edit
Unit: The unit for the oil flow can be changed
or
gal/h >
Cigarette smoke influences the measurement by more than
50 ppm. The breath of a smoker influences the
measurement by about 5 ppm.
When using an ambient CO probe, note that:
The direction of flow of the gas has an effect on the
accuracy of measurement. Frontal flow onto the probe
leads to higher measured values. The best measurement
results are achieved when the probe is moved gently
backwards and forwards.
When using the ambient CO probe and the flue gas probe,
note that:
the probe must be in the fresh air (CO-free) during the
zeroing phase!
→
]
Measurement options
→
→
[OK]
Oil Flow
Oil Flow
(of the oil nozzle) and
and partly [◄], [►]→ [OK].
clipboard: Any data saved to the clipboard
kg/h).
Spain
→
[OK]
6 Using the product
→ [OK].
Oil
→ [Edit].
[▼]
country version is
→
→
CO ambient
53