Oil Flow Calculation; Co Ambient - TESTO 300 Pro Smoke Edition Manual

Combustion analyzer
Table of Contents

Advertisement

Find Quality Products Online at:

9.6.10 Oil flow calculation

The function is only available if the chosen fuel is an oil.
This function is used to calculate the capacity of the burner from the set oil
pressure and the oil flow rate of the oil nozzle.
1
Call up function:
|
2
Set oil flow rate of the oil nozzle and oil pressure.
The calculated oil burner energy is displayed (in BTU/h).
Edit values
All values that can be modified have a dotted underlining.

9.6.11 CO Ambient

Cigarette smoke influences the measurement by more than 50 ppm.
The breath of a smoker influences the measurement by about 5
ppm.
When using an ambient CO probe, note that:
The direction of flow of the gas influences the accuracy of
measurement. Frontal flow onto the sensor leads to higher
readings. The best measurement results are achieved when the
probe is moved gently backwards and forwards.
When using the ambient CO probe and the flue gas probe, note
that:
The probe must be in the fresh air (CO-free) during the zeroing
phase.
An ambient CO probe (0632 1272) must be connected.
1
Call up function:
|
2
Start measurement: tap
Reading is displayed.
3
End measurement: tap
GlobalTestSupply
www.
9 Performing the measurement
Oil flow calculation
CO Ambient
.
.
53
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

300 pro300 smoke edition300Pro

Table of Contents