Before Next Use - Dräger Babytherm 8000 LDR Instructions For Use Manual

Warming bed
Table of Contents

Advertisement

Cleaning/Disinfecting/Sterilizing

Before Next Use

34
Caution: Do not use alcohol on the side and end panels; it will damage
materials.
Caution: Do not launder mattress in an automatic washing machine.
Do not autoclave mattress.
1. Follow manufacturer's instructions for use of the cleaning/disinfection
detergent for effective wipe-disinfection.
2. Wipe-disinfect the following surfaces:
side and end panels, inside and outside
resting surface
acrylic canopy
gel mattress (foam rubber mattress)
wooden bassinet, including attachments.
3. Remove visible impurities with a cloth and detergent.
4. Wipe with a damp cloth.
5. Dry with a clean cloth.
1. Reassemble components.
2. Operate unit for a few hours without an infant, in order to eliminate any
disinfectant residues from prior disinfection (at mattress temperature
setting of 37° C).
3. Check the wood surface for sharp nicks. Any sharp nicks on the surface
of the wood bassinet need to be repaired by authorized service
personnel.
Before an infant is placed in the Babytherm 8000 LDR:
4. Set up accessories for therapy as needed.
5. Check readiness for operation. See "Tests of Readiness for Operation" on
page 13.
6. Verify that mattress is sufficiently preheated.

Advertisement