tecan M1000 Operating Manual Addendum page 52

For the advanced operator
Table of Contents

Advertisement

48
9
When the test is completed, a PosID2 Test passed or PosID2 Test failed
message is displayed.
10 Confirm with <OK>.
11 Next select Close. The PosID2 Test Configuration screen is now displayed.
12 Select Print to print the test results.
13 Select Close to exit
Pass/Fail Criteria
If the PosID Test fails:
1
Check the positions of the barcode labels.
2
Clean the laser beam output window of the PosID. Refer to the m1000™
Operating Manual, Section 7.4.4.
3
Repeat the PosID test. If the test fails again, contact your local Abbott
representative.
The operator is prompted that the LiHa will move to the Home position. Select
<OK>.