Calibrate The Liha - tecan M1000 Operating Manual Addendum

For the advanced operator
Table of Contents

Advertisement

Calibrate the
LiHa
m1000™ Operating Manual Addendum, 392813, en, V1.1
Fig. 2
Teach worktable locations
6
Select Connect to the Instrument.
7
The operator is prompted to remove the sample and control racks, reagent
trough carriers, 1 ml subsystem carrier, DiTi racks, waste station chute, and
1.5 ml output tube carriers. Select <OK>.
8
Once the instrument is connected, the Connect to the Instrument button will
display Disconnect from the instrument.
9
Close the Safety door.
10 Select Initialise all Devices. The LiHa, RoMa, and PosID will move to the
Home position and the Instrument status button will change to Instrument
Status In Init during initialisation.
11 Once Initialization is complete, the Instrument status button will change to
Instrument Status Connected.
Calibrating the 1 ml subsystem for the LiHa
During this step, the LiHa is calibrated to correctly aspirate and dispense
specimens, controls, and reagents into the reaction vessels. This calibration will
include X, Y, and Z direction.
Connect
to the
Instrument
11