Lower Diti Eject Test - tecan M1000 Operating Manual Addendum

For the advanced operator
Table of Contents

Advertisement

Lower DiTi
Eject Test
m1000™ Operating Manual Addendum, 392813, en, V1.1
14 During the test, the results continuously appear on the Liquid Detector Test
Results window.
15 After the test is finished, the Liquid Detector Test passed or Liquid
Detector Test failed message is displayed. Confirm with <OK>.
16 Select Close.
17 Print the result with Print and/or finish the test with Close.
Pass/Fail Criteria
The Liquid Detection Test fails:
Check if tips are the same height.
Check DiTi cones are tight and clean. Refer to the m1000™ Operating
Manual, Section 7.3.2.
Check the liquid system for air bubbles. Refer to the m1000™ Operating
Manual, Section 7.3.2.
Repeat the Liquid Detection Test. If the test fails again, contact your local Abbott
representative.
The Lower DiTi Eject Test function tests the disposal of the disposable tips.
1
Start the GENESIS Setup & Service software by clicking on the
TECAN icon.
2
From the GENESIS Setup & Service main screen, select Liquid System
from the menu bar.
43