tecan M1000 Operating Manual Addendum page 35

For the advanced operator
Table of Contents

Advertisement

Selection
Flush
Flush tips once sequentially
The operator is prompted Do you want to
move LiHa above Washstation?
m1000™ Operating Manual Addendum, 392813, en, V1.1
5
Complete the following fields:
Field
Grid
Position
Use Fast Wash Pump
6
Close the Safety door.
7
Select one of the following functions:
8
Observe the dilutors for movement and leaks
If the dilutors are leaking, check the plunger screw and syringe screw for
tightness. Also check the valve connections for tightness. Refer to the
m1000™ Operating Manual, Section 7.3.2.
Change the syringes and/or 3-way valve. Refer to the m1000™ Operating
Manual, Section 7.5.
Repeat the Dilutor tool.
If the leaks continue, notify your local Abbott representative.
9
Select Print to print the report. Select <OK> to exit.
Selection
18
Waste
Function
Initializes all 8 dilutors at one time.
Initializes the dilutor one by one.
Select Yes.
31