Liquid
Detection Test
40
Corrective Actions
1
Check that the tips of the DiTi adapters are clean and are free of any
blockage. Refer to the m1000™ Operating Manual, Section 7.3.2.
2
Exit the GENESIS Setup & Service. Start NAT Wizard and perform a Flush
System and look for air bubbles. Refer to the m1000™ Operating Manual,
Section 7.3.6.
3
Repeat the Fast Wash Pump test. If the test fails again, contact your local
Abbott representative.
The Liquid Detection Test checks whether the liquid detection feature is working
properly.
1
Start the GENESIS Setup & Service software by clicking on the
TECAN icon.
2
From the GENESIS Setup & Service main screen, select Liquid System
from the menu bar.
3
Select Liquid Detection from the Liquid System menu and next select
Normal Test.
4
From the Select Worktable screen, select NAT MBS Worktable and then
<OK>.
NAT MBS
Worktable
m1000™ Operating Manual Addendum, 392813, en, V1.1