Select the following options:
Door Lock 1 (left)
Door Lock 2 (right)
9
Select Apply to save the configuration selected.
10 Close the safety door and check the door lock function by selecting Lock and
Unlock for the door lock selected.
11 If the door lock does not respond to the command, repeat the above step. If
the door lock still does not respond, contact your local Abbott representative.
12 Continue to check the door lock function by selecting Start from the
Access/Status Panel.
.
Start