Diagnostics: Status; Analyzer Status: Status: Logbook/Faults - Siemens ULTRAMAT 23 Manual

Continuous gas analysis gas analyzer for measuring ir-absorbing gases, oxygen and hydrogen sulfide
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

Functions
8.1 Diagnostics
8.1.1

Diagnostics: Status

8.1.1.1

Analyzer status: Status: Logbook/faults

104
In this menu you can call all status messages of the
ULTRAMAT 23 via further submenu items.
In this example, the status of the O
line. If the analyzer is equipped with software for operating an
H
S sensor, the text 'Probe status' is shown in the last line with
2
the following options:
• Only H
S sensor:
2
The analyzer calls this function directly.
• H
S and O
sensors:
2
2
You will be requested to select the associated sensor.
This dialog displays the contents of a logbook. This con-
tains all recorded faults. Each type of fault only appears
once in the logbook and is output in alphanumeric text (an
overview of the possible faults can be found in section
Faults (Page 178)).
Viewing the logbook is protected by code level 1.
Following access to the logbook you can:
• Display all recorded faults in succession using the <→>
key
• Delete the currently displayed fault message using the
<ENTER> key.
If further faults are present, these appear in succession.
You should delete all fault messages whose causes
have been eliminated.
Note
By deleting the fault message you do not eliminate the
cause of the fault (see section Faults (Page 178)).
A corresponding text is output when all stored faults have
been displayed. Terminate display of the logbook using the
<→> key.
Note
If the analyzer is in measuring mode, you can recognize
the occurrence of a fault in that an "F" appears at the right
edge. A "! at the right edge signals that a fault has been
logged which is no longer present.
sensor is shown in the last
2
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003

Advertisement

Table of Contents
loading

Table of Contents