Functions
8.4 Configuration
Configuration: Special functions: Factory data/reset/units: Reset
Configuration: Special functions: Factory data/reset/units: Change units
Note
The full-scale values may assume unusual values as a result of the component-specific
conversion factors. Subsequent adaptations can be carried out as described in
section Parameters: Measuring ranges: Setting measuring ranges (Page 126). In addition,
you should also check these parameters following this change:
• Setpoints of the calibration gases (section Calibration: Infrared measuring range: Set
span gas values (Page 113))
• Settings for the limits (section Parameters: Limits (Page 128)).
150
You can use this function to restart the analyzer.. When you
select this function, you will be warned (see
adjacent display) that the analyzer initially runs through a warm-
up phase following the restart and is thus unavailable for meas-
urements for a certain time.
The <ENTER> key initiates the restart with the warm-up phase.
Triggering of the restart can be prevented here using the
<ESC> key.
In the second line of this menu display, you can change the
factory-set units of the measured components.
After changing the unit, the display of the 'MB min' and 'MB
max' parameters is adapted accordingly.
This dialog is specific to the component.
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003