Gas inlet conditions
Sample gas pressure
Without pump
With pump
Sample gas flow
Sample gas temperature
Sample gas humidity
Note
Since measuring ranges can be changed, all accuracy data applies to the ranges specified
on the label!
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
Unpressurized (< 1200 hPa (17.4 psi) absolute)
Unpressurized suction mode, set in factory with 2 m (6 1/2 ft) hose
at sample gas outlet; full-scale value calibration necessary under
different venting conditions
72 ... 120 l/h (1.2 .. 2 l/min)
0 ... 50 °C (32 ... 122 °F)
< 90 % RH (relative humidity), non-condensing
Description
3.4 Technical specifications