installation

Make sure when mounting the analyzer that the environment is as free as possible of the
gas components to be measured!
In order to achieve the highest possible measuring quality, also observe the following
information concerning the location for mounting an analyzer!
Insufficient ventilation
The device may overheat or start burning in the case of insufficient ventilation.
• Ensure sufficient ventilation between the devices when installing in control cabinets. The
• Make sure during operation that the permissible ambient temperature range is always
NOTICE
Incorrect mounting
The device can be damaged, destroyed, or its functionality impaired through improper
mounting.
• Before installing ensure there is no visible damage to the device.
• Make sure that process connectors are clean, and suitable gaskets and glands are
• Mount the device using suitable tools. Refer to the information in Technical
Strong vibrations
Strong vibrations could loosen connections or damage sensors, resulting in free passage of
the sample gas into the environment.
Even weaker vibrations influence the result!
The analyzer must therefore only be used at a location which is free of vibration.
Please observe the data in the Technical specifications. (Page 31)
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
WARNING
heat sinks at the rear must remain free for air circulation.
observed (see Technical specifications (Page 31)).
used.
specifications (Page 31) for installation torque requirements.
CAUTION
4