Configuration: Special functions: ELAN/PROFIBUS/external interference: Cross-interference
8.4.2.4
Configuration: Special functions: Factory data/reset/units
Configuration: Special functions: Factory data/reset/units
Configuration: Special functions: Factory data/reset/units: Load factory data
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
The adjacent menu display appears when you select this
parameter. You can use this function to:
• Switch the correction of cross-interference on or off for the
duration of the calibration.
To do this, you must select the 2nd line and switch the
parameter on or off there.
– OFF (factory setting) means that the correction of
cross-interference is switched off during the calibration.
– ON means that the corrections of cross-interference
remain active during the calibrations. It is thus possible
to use gas combinations as calibration gases.
• Switch the correction of a constant cross-interference on
or off. To do this, you must select the 3rd line and switch
the parameter on or off there. If the constant cross-
interference is switched ON, the measured value of the
selected component is corrected with the entered value by
adding.
You can use this menu to select a number of items with which
e.g. inappropriate configurations and analyzer settings can be
cancelled:
The adjacent menu display appears when you select the func-
tion from the higher-level menu.
In this menu you can reestablish the original parameters pre-
sent when the analyzer was delivered.
Note:
All modifications (parameters and configuration) which you have
made since then are deleted.
The adjacent display appears when you select this function.
You define the further sequence by pressing either the
<ENTER> or <ESC> key.
When you select this function, the adjacent display appears for
the duration of the load procedure.
Functions
8.4 Configuration
149