Siemens ULTRAMAT 23 Manual page 175

Continuous gas analysis gas analyzer for measuring ir-absorbing gases, oxygen and hydrogen sulfide
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

Proceed as follows to install the new sensor:
1. Tighten the new H
Figure 10-2
2. Plug the cable with the connector onto the sensor.
3. Close the analyzer by screwing tight the cover.
4. Switch the analyzer on and wait for the warming-up phase.
5. Check for leaks in the device
This is described in section Leaks in the gas paths (Page 81)
6. Calibrate the H
7. Enter the installation date with the menu command "H
The analyzer is then ready for use again.
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
S sensor by hand in the holder up to the mechanical limit (shaded)
2
Sensor holder
(shaded)
S sensor as described in section Calibration: H2S sensor (Page 118)
2
Service and maintenance
10.2 Maintenance work
S Installation date".
2
175

Advertisement

Table of Contents
loading

Table of Contents