Service Calls

When calling technical support about your unit, be prepared to give the
serial number from the product identification label. When giving the
serial number, the technical support representative can identify your unit
and provide the information you need more quickly.

Error Messages

For warnings that may appear on the display window, refer to the table
below.
Warning
Er1
The measured value is
below the range of the
device. In the case of an aver-
aging measurement, the mea-
surement fluctuation is
excessively large.
The device is malfunctioning. Take the unit out of service,
Er2
A measurement error may
through
have occurred during an aver-
Er6
aging measurement, or the
hardware is not functioning
properly.
Jaundice Meter Instructions for Use (MU01380)
DRAFT 18 May 2005
Error Messages
Cause
Section 7
Troubleshooting
Solution
Take the measurement again.
If Er1 still appears, use
another device to repeat the
measurement, or perform a
serum bilirubin test.
and contact the nearest autho-
rized service facility.
Set the Power switch to the
Off position, and then return
it to the On position. If the
warning still appears, contact
the nearest authorized service
facility.