Dräger JM-103 Instructions Manual page 23

Jaundice meter
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration—Electromagnetic
The Jaundice Meter is intended for use in the electromagnetic envi-
ronments specified below. The customer or user of the unit should
ensure that the unit is used in such environments.
Immunity
Test
Electrostatic
discharge
(ESD)—
IEC 61000-
4-2
Electrical
fast tran-
sient/burst
—IEC
61000-4-4
Surge—IEC
61000-4-5
Jaundice Meter Instructions for Use (MU01380)
DRAFT 20 June 2005
Immunity
IEC 60601
Compliance
Test Level
± 6 kV con-
± 6 kV con-
tact
tact
± 8 kV air
± 8 kV air
± 2 kV for
± 2 kV for
power sup-
power sup-
ply lines
ply lines
± 1 kV
± 1 kV
differential
differential
mode
mode
Electromagnetic
Level
The floors should be
wood, concrete, or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Environment—
Guidance
Page 2 - 13