Guidance and Manufacturer's Declaration—Electromagnetic
The Jaundice Meter is intended for use in the electromagnetic environ-
ments specified below. The customer or user of the unit should ensure
that the unit is used in such environments.
Immunity
Test
Con-
ducted
RF—IEC
61000-4-6
Radiated
RF—IEC
61000-4-3
Jaundice Meter Instructions for Use (MU01380)
DRAFT 20 June 2005
IEC
Compli
60601
ance
Test
Level
Level
3 Vrms
3 Vrms Portable and mobile RF communica-
150
kHz to
80
MHz
3 V/m
3 V/m
80
MHz
to 2.5
GHz
Immunity
Electromagnetic Environment—
Recommended Separation Distance
tion equipment should be used no
closer to any part of the Jaundice
Meter, including cables, than the rec-
ommended separation distance calcu-
lated from the equation applicable to
the frequency of the transmitter.
Recommended Separation Distance
d
=
1.2 P
d
=
2.3 P
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended separa-
tion distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey
the compliance level in each frequency
range.
b
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
Guidance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
, should be less than
a
Page 2 - 15