Dräger JM-103 Instructions Manual page 55

Jaundice meter
Table of Contents

Advertisement

Symptom
The display is blank
when the Power switch
is in the On position.
The display suddenly
goes blank during a
measurement.
The Charger lamp
does not illuminate
when the Jaundice
Meter is placed on the
charger unit.
It is impossible to take
measurements.
Jaundice Meter Instructions for Use (MU01380)
DRAFT 18 May 2005
Troubleshooting
Possible Cause
The batteries are
exhausted.
The batteries are
exhausted.
The Jaundice Meter is
not placed in the
charger unit correctly.
The charger unit and
the AC adapter are not
plugged into an AC
outlet correctly.
The Ready lamp is not
illuminated.
The batteries are
exhausted.
Action
Charge the battery (see
"Charging the Battery"
on page 4-1).
Charge the battery (see
"Charging the Battery"
on page 4-1).
Place the Jaundice
Meter in the charger
unit aligned perpendic-
ular to the measuring
point with the display
facing forward.
Plug the charger unit
and the AC adapter into
an appropriate power
source correctly.
Before taking a
measurement, ensure
that the Ready lamp is
illuminated.
Charge the battery (see
"Charging the Battery"
on page 4-1).
Page 7 - 3