Guidance and Manufacturer's Declaration—Electromagnetic
The Jaundice Meter is intended for use in the electromagnetic envi-
ronments specified below. The customer or user of the unit should
ensure that the unit is used in such environments.
Immunity
Test
Voltage
dips, short
interrup-
tions, and
voltage vari-
ations on
power sup-
ply input
lines—IEC
61000-4-11
Power fre-
quency
(50/60 Hz)
magnetic
field—IEC
61000-4-8
NOTE:
U
is the AC mains voltage prior to the application of the test
T
level.
Page 2 - 14
DRAFT 20 June 2005
Immunity
IEC 60601
Compliance
Test Level
< 5% U
(>
< 5% U
T
95% dip in
95% dip in
U
) for 0.5
U
T
cycles
cycles
40% U
40% U
T
(60% dip in
(60% dip in
U
) for 5
U
T
cycles
cycles
70% U
70% U
T
(30% dip in
(30% dip in
U
) for 25
U
T
cycles
cycles
< 5% U
(>
< 5% U
T
95% dip in
95% dip in
U
) for 5
U
T
seconds
seconds
3 A/m
3 A/m
Jaundice Meter Instructions for Use (MU01380)
Level
(>
Mains power quality
T
should be that of a typical
commercial or hospital
) for 0.5
T
environment. If the user
of the unit requires con-
T
tinued operation during
power mains interrup-
) for 5
T
tions, it is recommended
that the unit be powered
T
from an uninterruptable
power supply or battery.
) for 25
T
(>
T
) for 5
T
The power frequency
magnetic fields should be
at levels characteristic of
a typical location in a typ-
ical commercial or hospi-
tal environment.
Electromagnetic
Environment—
Guidance