CAUTION:
Unit must be placed in charger as shown, or damage to charging
contacts could occur.
3.
Place the Jaundice Meter on the
charger unit so that its display faces
you. When the Jaundice Meter is set
on the charger unit properly, the
Charger lamp lights up.
NOTE:
With a fully charged battery, approximately
400 measurements can be taken.
4.
Allow approximately 32 hours for
charging to complete.
CAUTION:
The instrument has a built-in, non-user-replaceable battery. Do
not disassemble the instrument to replace the battery. To replace
the battery, contact your dealer or authorized service center.
Failure to do so could result in equipment damage.
5.
To replace the battery, contact your dealer or an authorized service
facility.
Page 4 - 2
DRAFT 20 June 2005
Jaundice Meter Instructions for Use (MU01380)