Specifications

Standard Features

Feature
Model name
Measuring method
Measurement range
Clinical Data Standard Error of
Estimate (SEE) *
Light source
Light source life
Detectors
Power source
Protection type and level
Minimum number of
measurements when fully charged
Operating temperature range
Operating relative humidity range
Storage temperature range
Storage relative humidity range
Dimensions
Weight, including Ni-MH battery
AC adapter input
*The standard deviation shown above is based on the average of the
clinical data available. On average, 66% of results fall within this range,
and the remainder fall outside this range. This value can be affected by
variables such as age, skin color, and preformance of the device in the
hands of the user. Refer to Appendixes A and B for a detailed
description of results by clinical site, measurement location, and patient
demographics. The SEE shown in the table are based on the clinical data
Page 2 - 10
DRAFT 20 June 2005
JM-103
Determines the yellowness of the
subcutaneous tissue by using two
optical paths to measure the optical
density difference at two wave-
lengths
0.0 mg/dL to 20 mg/dL or 0 μmol/L
to 340 μmol/L
± 1.5 mg/dL or ± 25.5 μmol/L
Pulse xenon arc lamp
150000 measurements
Silicon photodiodes
2.4 V, Special Ni-MH battery
Internally-powered instrument, BF-
type
400 single measurements
10°C (50°F) to 40°C (104°F)
30% to 95% non-condensing
-10°C (14°F) to 50°C (122°F)
30% to 95% non-condensing
4.8 cm (1.9") wide x 15.4 cm (6.0")
high x 3.2 cm (1.2") deep
150 g (5.3 oz)
100V - 240V 50/60Hz, 11-18VA