Siemens SIMATIC HMI OP 73micro Operating Instructions Manual page 99

Wincc flexible
Table of Contents

Advertisement

Procedure - restore
Proceed as follows:
1. When restoring with reset to factory setting only:
2. From the Windows Start menu, run ProSave on the PC.
3. Select the HMI device type from the "General" tab.
4. Set the connection parameters in the "General" tab.
5. Select the *.psb backup file to be restored from the "Restore" tab.
6. To restore the password list:
7. Start the restore operation in ProSave with "Start Restore".
Result
The transfer is completed when the backup data is restored from the PC to the HMI device.
See also
Configure the data channel (Page 88)
Configure the data channel (Page 79)
Setting the operating mode (Page 90)
Overview of backup and restoring (Page 95)
Data Transfer Options (Page 92)
OP 73micro, TP 177micro (WinCC flexible)
Operating Instructions, 09/2007, 6AV6691-1DF01-0AB0
Switch off power to the HMI device.
The tool outputs a message showing the HMI device for which the backup copy was
created and the type of backup data contained in this file.
Set "Transfer" mode on the HMI device.
Follow the instructions in ProSave.
A status view opens to indicate the progress of the operation.
Preparing and backing up a project
7.3 Backup and restore