14.0 Summary Report Test - ZOLL E Series Service Manual

Defibrillator
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

14.0 Summary Report Test

Tools Needed: QED 6 Defibrillator Analyzer.
Test Setup:
1. Connect the universal cable to the QED 6 analyzer.
2.
If you are using paddles, place the paddles on the analyzer's discharge plates.
Do this...
14.1
Press and hold the SUMMARY softkey for 4 to
8 seconds to display erase options. Press the
ERASE SUMMARY softkey to erase any
previously stored data erase any previously
stored data.
14.2
Set selector switch to DEFIB. Select 200J using
the ENERGY SELECT button, and press the
CHARGE button. When charged, press the
SHOCK button to discharge into the defibrillator
analyzer.
14.3
Wait 18 seconds, then press the Code Marker
softkey. Press the CPR softkey.
14.4
Turn the unit off. Wait 10 seconds and then turn
the unit on. Press the SUMMARY softkey, then
press the PRINT CHART softkey.
Record your results on the Maintenance Tests Checklist.
9650-1450-01 Rev. D
Verify that...
The unit displays the message, ERASING REPORT.
The unit successfully discharges and prints a strip chart.
The Code Markers appear.
The summary report prints. The report displays the correct date, time, the
shock delivered and Code Marker event.
E Series Service Manual
Pass/Fail
o o
o o
o o
o o
1-25