Recorder

Symptom
Unit displays the CHECK RECORDER prompt.
Recorder makes stuttering sound when activated.
Light or poor quality printing is observed.
Summary report does not print when you press
SUMMARY button.
9650-1210-01 Rev. R
Recommended Action
Ensure the recorder door is shut properly.
Ensure adequate paper supply in recorder paper tray
(see "Changing Recorder Paper" on page 11-4).
Remove paper, check paper type, check recorder for
paper jam, and then reload paper.
Check recorder for paper jam.
Ensure correct paper type (ZOLL P/N 8000-0300) is in
use.
Ensure paper is installed grid-side against recorder print
head.
Ensure that door is latched shut properly. You must
listen for clicks on both corners of the door when
latching the door shut.
Clean recorder print head (see "Cleaning the Recorder
Printhead" on page 11-1).
Wait 15 seconds and try again. (The unit requires 15
seconds after an event occurs to finish storing a record
of the event. A summary report cannot be printed during
this time.)
Troubleshooting Guidelines
13-3