ZOLL E Series Service Manual page 50

Defibrillator
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

E Series Service Manual
Do this...
23.1
Turn the selector switch to MONITOR mode.
(For AED units, turn the selector switch to ON
and select Manual mode.)
23.2
Ensure that the LEADS parameter is set to
PADS (default).
If necessary, press the LEADS button to cycle
through the values to select PADS.
23.3
Press the NIBP button on the E Series front
panel.
23.4
Press the PARAM softkey, then press the
TREND softkey, followed by the NIBP Trend
softkey.
Record your results on the Maintenance Tests Checklist.
4
These values only apply for test set-ups utilizing the BP Pump 2 Simulator. Variations of the test set-up or different simulators may produce readings outside
the provided values and will require end-user facility to establish the appropriate offset and tolerances. Please contact ZOLL Technical Support i f you require
assistance establishing the offset o f y o u r simulator and test set-up.
1-40
Verify that...
The E Series unit displays PADS in the Lead selection field on the monitor.
The E Series displays the following measurements:
• Systolic pressure (115 - 125 mmHg)
• Diastolic pressure (75 - 85 mmHg)
• Mean pressure (88 - 98 mmHg)
The E Series unit displays a summary of the NIBP measurements, including
the pulse rate reading (in the range of 77 - 83 bpm).
Pass/Fail
o o
o o
o o
9650-1450-01 Rev. D