Capturing A Data Snapshot; Reviewing And Printing Snapshots; Treatment Summary Report - ZOLL Propaq MD Operator's Manual

Portable defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
16 P
HAPTER
ATIENT

Capturing a Data Snapshot

Press the Snapshot button (
waveform patient data. The unit captures 12 seconds proceeding and 12 seconds following the
button press.
The Propaq MD unit can store a minimum of 32 snapshots, including
Monitor snapshots
Defibrillator snapshots
Treatment snapshots
Pacer snapshots
Alarm snapshots

Reviewing and printing snapshots

You can perform this procedure from any main display screen.
1. Press the Home/Display button (
Use the navigation keys to get to the Trends window.
2.
Press Select button to scroll through the list of trends.
3.
Note:
Select the desired snapshot from the Trends list and press Select. The snapshot numeric data
4.
displays.
To print the data and the snapshot waveforms, highlight Print this Snapshot and press
5.
Select.

Treatment Summary Report

The Treatment Summary Report is a report that shows all of the treatment events related to a
patient such as defibrillation/pacer events, alarm events, and treatment snapshots. It is helpful
to print out this report at the end of a case.
TREATMENT SUMARY REPORT
Name: John Doe
ID: Patient 0015
Start Time:
Last Event: 09/14/10
Elapsed Time: 00:29:48
Total Shocks: 1
Total Pace Time: 00:00:00
Total 12 Leads:
Dept: ICU
Unit:
S/N: 0
SW: 00.00.00.00
16–2
D
ATA
) on the front panel to capture a 24-second period of numeric and
The snapshots have a snapshot icon next to the time stamp.
Patient Mode: Adult
09/14/10
09:14:14
09:43:40
# Events:
0
www.zoll.com
) repeatedly until the Trends window displays.
09:14:14
09:14:14
09:14:14
09:14:14
09:14:20
09:14:43
New Case ID
System On
Patient Mode Adult
Some Alarm Limits Disabled
Self Test Passed
Treatment Snapshot: ASA
9650-0802-06 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents