Heart Rate Alarm Activated Format

The summary report function records 6 seconds of pre-alarm patient ECG. Also recorded are the ECG lead, ECG size,
patient's heart rate, actual event time, and date. The date/time printed on the top of the strip corresponds to the ECG
data occurring 6 seconds before the event. If the pacer is on during this event, the pacing rate and pacing current are
also recorded.
Actual
event
date/time

VF Alarm Activated Format (Refer to Section 6)

The summary report function records 18 seconds of patient ECG data associated with each VF alarm. Also recorded
are the shock count, ECG lead, ECG size, actual event time, patient's heart rate, and noise events. The date/time
printed on the top of the strip corresponds to the ECG data occurring 6 seconds before the event.
Actual
event
date/time

Recorder On Format

The summary report function records 6 seconds of patient ECG prior to turning on the recorder. Also recorded are the
ECG lead, ECG size, patient's heart rate, actual event time, and date. The date/time printed on the top of the strip
corresponds to the ECG data occurring 6 seconds before the event. If the pacer is on during this event, the pacing rate
9650-1210-01 Rev. R
Operating Controls and Indicators
2-7