Connector Panel Removal

Note:
There are 2 styles of SPO2 connector (MS-11 & MX-1) used on the E Series, use the appropriate instructions for disassembly.
1. (MS-11 Style) Remove the 2 screws securing the panel to the
housing.
3. Remove rubber gasket (black foam) from connector panel. ECG
and SPO2 connectors can now be removed.
Note:
During reassembly ensure conductive sleeve is in place next to SPO2
connector.
9650-1450-01 Rev. D
2. Lift assembly out, gently sliding rubber ground sleeve off
mounting post.
E Series Service Manual