Recommended Separation Distances From Rf Equipment For E Series Non-Life-Support Functions - ZOLL E Series Service Manual

Defibrillator
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

E Series Operator's Guide
Recommended Separation Distances from RF Equipment for E Series
Non-Life-Support Functions
The non-life-support functions
disturbances are controlled. You can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the E Series unit as recommended
below, according to the maximum output power of the communications equipment.
Rated Maximum Output Power of
Transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
The non-life-support functions on the E Series unit are defined to be any function not listed as a life-support function
in the "EID for Life Support Functions" table (Note a). Specifically, these functions are noninvasive blood pressure
(NIBP), end-tidal CO
(EtCO
2
A-12
a
of the E Series unit is intended for use in an environment in which radiated RF
Separation Distance According to Frequency of Transmitter
150 kHz to 80 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
), and SpO
.
2
2
(m)
80 MHz to 800 MHz
P
d = 0.35
0.035
0.111
0.35
1.11
3.5
800 MHz to 2.5 GHz
P
P
d = 0.7
0.07
0.22
0.7
2.2
7.0
9650-1210-01 Rev. R

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents