E Series Service Manual
Error Message
REPLACE BATTERY
REPLACE CARD
REPORT FULL
REPORT HALTED
SELECT 30J FOR TEST
SELECT DEFIB MODE
SELECT LEADS
SELECT LIMB LEADS
SELECT PADS
SET CLOCK
SET PACE mA
SET PACE RATE
SHOCK ADVISED
SpO
AMBIENT LIGHT
2
2-12
Explanation
Battery voltage is less than absolute minimum. Shutdown
imminent.
Write errors during manual or semi-automated modes.
Summary report memory full.
Summary report stops printing unexpectedly.
Attempt to run a self test at an energy other than 30J.
Analyze button pressed in pace or monitor mode.
12 Lead monitor determined the user selected MFE.
Paddles or augmented ECG leads selected when
continuous analysis active or started.
Lead I, II, or III selected when analyze pressed.
Real time clock failure: invalid date or time.
Multiple copy errors are the product of intended software
or memory errors. If error reoccurs other than on entering
pace the first time or after more than 10 minutes in other
mode, the unit could be broken.
Multiple copy errors are the product of intended software
or memory errors. Multiple copies of pace rate don't
match. If error persists, unit could be broken
Advisory message when analysis finds a shockable
rhythm.
Ambient light is too bright.
User
Technical Action
Advisory
Replace with charged battery.
May have configuration card installed or write
protection on.
Erase summary report.
Turn unit off and then back on again. Print
Summary again. If fault persists, replace system
board.
Select limb leads I, II, III or MFE
Set date and time and/or replace system board.
Set pace current. If broken, replace system board.
Set pace rate. If broken, replace system board.
Shield sensor from ambient light. Replace Sp0
sensor. Replace Sp0
2
2
module
9650-1450-01 Rev. D