ZOLL E Series Service Manual page 41

Defibrillator
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

18.0 EtCO
Monitor Test (for EtCO
2
Tools Needed: CAPNOSTAT 5 Mainstream cable with airway adapter, or CAPNOSTAT 5 Sidestream cable with cannula.
Test Setup:
1. Install the battery.
Do this...
18.1
Connect the CAPNOSTAT 5 CO2 Mainstream
cable with airway adapter attached, or the
CAPNOSTAT 5 Sidestream cable with the
cannula attached, to the yellow connector at the
back of the E Series.
18.2
Set the front panel switch to MONITOR or ON.
For AED units, enter Manual Mode.
18.3
When the WARM UP message disappears,
press the Param softkey, then select EtCO2
and press Enter.
Press the ZERO softkey, then wait for the
ZERO DONE message to appear.
Press the Return softkey, then press the 
18.4
Wave 2 softkey until the CO
appears.
18.5
Breath normally into the airway adapter.
Record your results on the Maintenance Tests Checklist.
9650-1450-01 Rev. D
Option)
2
Verify that...
The airway adapter (for Mainstream) or the cannula (for Sidestream) is
installed in the CO
The E Series monitor displays the WARM UP message.
Note: Warm up may take about 1 minute.
The E Series monitor displays the ZERO DONE message.
A flat baseline CO
waveform
2
A capnogram waveform appears.
cable.
2
waveform appears.
2
E Series Service Manual
Pass/Fail
o o
o o
o o
o o
1-31