17. Pacer Output mA (Pacer Version Only)
When pacing is selected, this control sets the amount of
current delivered to the MFE Pads. For conscious
patients, it should be gradually increased until capture is
recognized. The selected current setting is indicated on
the display.
18. 4:1 BUTTON (Pacer Version Only)
Tests for threshold or to determine the patient's
underlying rhythm. When depressed, this button causes
pacing stimuli to be delivered at ¼ of the frequency of the
current ppm setting. The device resumes normal pacing
operation when the button is released.
19. Pacer Rate ppm (Pacer Version Only)
When pacing is selected, this control sets the rate at
which the pacemaker will operate. It must be set above
the patient's intrinsic rate in order for the pacemaker to
provide stimulation. The selected pace rate setting is
indicated on the display.
20. Systole and Alarm Speaker
Emits an audible heart rate tone during ECG monitoring,
and audible alarm indications when an alarm condition
occurs.
21. Microphone (Optional)
Records audio activity in the vicinity of the E Series unit
for storage on the PCMCIA data card.
22. NIBP Button (Optional)
Allows you to start single, auto, or STAT noninvasive
blood pressure measurements as described in the option
insert Non-Invasive Blood Pressure (part number
9650-1214-01). Your unit has this button only if you
ordered this configuration.
9650-1210-01 Rev. R
Operating Controls and Indicators
The next three items are located on the top of all units,
as shown in the following figure.
Data Card Slot
Stripchart
Recorder
Compartment
Stripchart Recorder Compartment
Located on top of the unit, the recorder compartment
holds the paper supply for the recorder. Open the cover
to replace the paper.
PCMCIA Data Card Slot
Located on top of the unit, the PCMCIA data card slot
holds the PCMCIA flash memory card for data storage
and retrieval.
PC Card Modem Slot (12-Lead Option Only)
Located on top of the unit, the PC Card modem slot
holds the modem card for transmitting 12-Lead ECG
information to remote locations via landline or cellular
phone. See the 12-Lead ECG Monitoring insert (part
number 9650-1213-01) for more information.
Charge Indicator Light (Not Shown)
Located on the apex paddle, this light turns on when the
defibrillator is charged and ready.
Defibrillator Test Port (Not Shown)
Located on the Multi-Function Cable, the test connector
is used to test the defibrillator output using the
Multi-Function Cable only.
Pediatric Paddles
Pediatric-size electrodes are built into the paddle
assembly; they lie directly under the adult electrode
surface. Access them by pushing the black PEDI button
at the front of each paddle and sliding the adult surface
forward. When replacing the adult electrodes, it is
PC Card
Modem Slot
2-3